Attributes
Key | Value |
---|
Auto ? OFF | Yes |
Auto-Check | Self-test at each switc. |
Automatic Cuvette Recog. | Round cuvettes: 13, 16 . |
Compliance | CE |
Dimensions | 422 x 195 x 323 mm |
Display | 7? high contrast colour. |
Drift | < 0.005 Abs per hour af. |
Interference Emission | Class B |
Interference Immunity | IEC 61000-4-3 |
Internal Storage | approx. 5000 data sets . |
IP Protection Class | EN 60529 |
Languages Full User Man. | German, English, Spanis. |
Languages Quick Start G. | German, English, French. |
Languages User Interface | German, English, French. |
Light Source | Tungsten halogen lamp |
LIMS Compatibility | ASCII, .csv-files |
Measurement | Concentration, single a. |
Meter Safety | EC Directive 2014/35/EC. |
Operation | Membrane Keyboard |
Optics | Grid monochromator with. |
Photometric Accuracy | 0.003 Abs below 0.6 Abs. |
Photometric Linearity | < 1 % up to 2.0 Abs bet. |
Photometric Range | -3.3 - +3.3 Abs |
Photometric Reproducabi. | 0.003 Abs below 0.6 Abs. |
Photometric Resolution | Absorption: 0.001 ; Tra. |
Portability | Benchtop |
Power Drain | 100 - 240 V, 50/60 Hz |
Power Supply | Buffer batteries (4 x A. |
Protection Class | IP 30 |
Scan Speed | 700 - 2000 nm/min. |
Security | Password protection pos. |
Spectral Scope | 4 nm |
Stock Conditions | -25 ?C to +65 ?C (-13 ?. |
Stray Light | > 0.1 % Transmission at. |
Test Recognition | via internal barcode re. |
Tolerance Extension | 0.008 E |
Wavelength Accuracy | ? 1 nm on all Holmium p. |
Wavelength Range | 320 - 1100 nm |
Wavelength Reproducibil. | better than 0,5 nm |
Wavelength Resolution | 1 nm |
Weight | 4.5 kg |